Communication Menu
Activate sensor command - output
Activate sensor command - input

9.3 USB-A port (host)

This option allows the export or import of data from an external mass storage device. The
device must first be connected to the instrument through the USB-A port.
Select the export option to export data from the instrument to the storage device. This will export
all stored data: measurements, calibrations, product list, access table, etc.
For both import and export options, the progress bar is updated to give an indication of the
progress of the selected option.
The USB-A port is also used to update the instrument software. If the software is present on the
USB storage device, the update process starts automatically.
82
Name
Value
OCT
1-2
OCI
2
Channel number:
OCD1
0
0 = Channel 1
Sensor activation:
OCD2
0-1
0 = Deactivate the EC sensor
1 = Activate the EC sensor
OCD3
Not used
OCD4
Not used
Name
Value
ICT
1-2
0 = OK
1 = Unknown command ID
ICS
0-3
2 = Invalid parameter (e.g. invalid channel no)
3 = Execution failure
ICD1
Not used
ICD2
Not used
ICD3
Not used
ICD4
Not used
Comment
Comment
Select one of the two import options (product list or access
table) to import data from the storage device. This is useful for
transferring these files to additional instruments without the
need of having to re-enter the data individually on each
instrument.
Note: The imported data will override any current settings on the
instrument.