9.4 HTTP/TCP-IP

9.4.1 Overview

When activated this option downloads data from the instrument directly to a web page that can
be accessed from a PC. To be able to use this option, the instrument must be connected to the
network (specifically Connector P3. Refer to
must have a DHCP server installed.

9.4.2 PC interface

Once the server has been enabled and the interface information set up, access the information
by launching an internet browser and typing "http://" followed by the server name that has been
assigned to the instrument, in the address box as illustrated below:
Main board connections on page
Check the Enable server box to enable the web server
communication link.
Enter the Server name for the instrument. This is free
format text and should typically be used to identify the
instrument.
Check the Enable security box if you require a password
to be entered on the PC to access the web page.
If any of the details on the previous screen have been
changed, a warning message will be displayed as illustrated
left.
The changes must be confirmed, after which the instrument
must be powered down and powered up again for the
changes to take effect.
Note: The IP Diagnostics button at the bottom of the screen is for use
by experienced IT personnel only to help resolve any
communications problems.
If the enable security option has been checked on the
instrument, you will be required to enter a username
and password on your PC to gain access to the web
page.
The username and password must be a valid
username/password combination that has been set up
on the instrument (refer to
page 86
for user set up information). Domain
information is not required.
Communication Menu
33. The network
User management on