Installation
3.7.3 Measurement board
Connector J7 (inputs & outputs)
Measurement alarms relays
1. Common
2. Output relay 1
3. Output relay 2
4. Output relay 3
Digital inputs
9. to 16. Not used
3.8 Measurement alarm relays
Figure 18 Output Relays
34
Figure 16 Measurement board
Analog current (or voltage) outputs
5. GND
6. Output 1
7. Output 2
8. Output 3
The three output relays are located on the measurement
board.
They can be individually configured to Normally Open (NO) or
to Normally Closed (NC) by physically moving the jumper on
each relay. On the picture example on the left:
Upper relay is set to NC
•
Middle relay is set to NO
•
Lower relay is shown with no jumper, to show the 3 pins.
•
Note: J4 (upper relay) is relay 1
J5 (middle relay) is relay 2
J6 (lower relay) is relay 3
Figure 17 Connector J7