Services Menu

13.2 Language selection

13.3 Clock

13.4 Screen

13.4.1 Screen calibration

13.4.2 Screen brightness

94
Note: If Zero diagnostic is selected, the warning message illustrated
left appears. Ensure the sensor has been calibrated in pure N
selecting this option.
Check the language as required and restart the instrument to
apply the change. The instrument will restart in the language
selected
Note: Level 3 or 4 is required to change the display language.
Type in each appropriate box the actual time and date, and
select the display format for them.
This Windows CE screen allows you to adjust the click
position corresponding to the displayed buttons. Use it if ever
the sensitive areas are no longer properly aligned with the
buttons on display. Follow the instruction given on screen:
Place the stylus right on the cross when asked and proceed.
User will be asked to click on the screen to accept the new
setting. If not, the new setting is not recorded and no change
is made.
Press the up or down arrow to increase or decrease the
screen brightness. Four levels are pre-set: Max,
Comfortable, Standard (default setting) and Economy.
Press OK when finished.
Note: This can also be called through the brightness icon on
the main display.