Sensor Spot Replacement; Storage, Handling And Transportation - Hach ORBISPHERE G1100 User Manual

Sensor and analyzer
Table of Contents

Advertisement

Maintenance and Troubleshooting

14.2.3 Sensor spot replacement

1. Take the new sensor spot and place into the end of
2. Using your fingers, put the new securing O-ring in
3. To ensure the O-ring is properly in place, take the

14.3 Storage, handling and transportation

Protect the instrument against the elements: rain, splashing, direct sunlight, etc.
A properly packaged instrument can be stored and transported at a temperature -20°C to +70°C
and relative humidity up to 80%. Best practice for packing the instrument for transportation is to
reuse the original packaging in which the instrument was first delivered. The instrument should
be stored in suitable premises, free of dust, condensation and chemical evaporation.
In cold weather, avoid sudden temperature change (like when entering a warm room) and give
the instrument enough time to adapt to the ambient temperature in order to avoid condensation
inside.
To clean the instrument, wipe the housing clean with a cotton cloth or tissue. Always clean the
instrument before storage. Pay attention not to scratch the surface of the display to retain good
clarity over time.
Never use liquids such as oil, benzene, or detergents for cleaning the instrument. A mild glass
cleaner can be used to remove greasy stains.
98
the sensor using the maintenance tool.
place and push into position.
maintenance tool and put the large end on top of
the O-ring on the sensor. Push down firmly and
twist left and right a few times until the O-ring is
flush with the top of the sensor.
NOTICE