Advertisement
SAFETY NOTE
Chemical exposure hazard. Refer to the current material safety data sheets (MSDS) for safety protocols.
The H135 miniLab is a portable waterproof pH meter with an ISFET pH sensor for pH measurements for the food and beverage industry.
Calibration points | 1, 2 or 3 point |
Auto buffer recognition | pH 4.01, 6.86, 7.00, 9.18, 10.01 |
Temperature compensation | Automatic |
pH range | 2 to 12 pH |
Temperature range | 5 to 40°C (40 to 105°F) |
Power | Auto Shut Off. 3 V lithium batteries (2x), CR2032 |
To change the buffer set or the temperature units, push and hold CAL 1 and then push the power button. Push CAL 1 again to toggle between the parameters.
For protein deposits the manufacturer recommends an enzyme cleaning solution (2964349). Do not use organic solvents.
Always dry the meter first.
Avoid direct contact with the pH sensor.
When the liquid in the reference electrode is decreased, replace the electrode (9316900).
Error code | Definition |
Battery icon | <10 hours battery life |
BAT | Replace the battery. |
0.00 or 14.00 | Make sure that the probe is properly immersed in the sample. |
E06 | Slope error. |
E07 | Voltage error. |
E08 | Calibration time out. |
E13 | Temperature sensor error. |
Note: When an error occurs, clean the probe and calibrate again. If the problem is not solved, contact Technical Support.