Hach 50 14 T Probe Manual

Introduction

This pH refillable probe with temprature sensor is used in general water samples, samples with sulfides and samples with reduction substances.

50 14
Refer to the meter manual for more information.


caution Personal injury hazard. Glass components can break. Handle with care to prevent cuts.
caution Chemical exposure hazard. Refer to the current material safety data sheets (MSDS) for safety protocols.

Technical information

Range 0 to 14 pH
Junction Ceramic (2)
Electrolyte KCl, 3 M
Reference Ag ion barrier cartridge
Temperature Pt 1000
Minimum depth 20 mm (0.8 in.)
Operating temp –10 to 100°C (14 to 212°F)
Limits Viscous solutions, distilled water,
samples with colloids

Connect the probe

Connect the probe Calibration

Calibration

Calibration - Step 1
caution Air bubbles in or under the probe tip when submerged can cause slow stabilization or error in measurement.
Calibration - Step 2

Measurement

Measurement

Maintenance and storage

Cleaning: soak the probe in a cleaning solution. Rinse with deionized water.

Contamination Cleaning agent
Proteins Pepsin cleaning solution
Grease, oils, fats Electrode cleaning solution
Limescale 0.1 N HCl solution

Hydration: soak the dry probe in a dilute HCl solution for several hours.
Storage: put the probe in the protector with KCl 3 M solution.

Troubleshooting

Be sure that the samples are suitable for the application.

Warranty

1 year for manufacturing faults only. Damage from use is not covered.

Documents / Resources

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Hach 50 14 T Probe Manual

Advertisement

Need Assistance?

Do you have a question about the 50 14 T Probe that isn't answered in the manual? Leave your question here.

Error while submitting your question. Please, reload page and try again.

Thank you! Your question has been received!

Table of Contents