Installing the Temperature Stabilizer on the AM

1. Remove the plate on the right side of the AM that covers the temperature
stabilizer mounting location.
2. Slide the temperature stabilizer into the opening and tighten the captive
thumbscrews (see
B.4.5

Installing the AM in the DC Upper Compartment

Always turn off the DC power before installing an AM.
1. If the DC power is on, press the
2 seconds to turn off the power before continuing.
2. Locate the high-pressure valve labels shipped with the AM.
3. Unscrew the black locking ring on the front of the high-pressure valve.
Attach the appropriate label (6-port or 10-port) to the front of the AM.
Replace the black locking ring.
4. Pull the DC top component tray all the way forward (see
Figure B-28. DC with Top Component Tray Pulled Forward
5. Locate the pair of mounting brackets (P/N 071260) shipped with the AM.
150
Figure
B-24).
POWER
button on the front of the DC for
Figure
Openings for AM
Mounting Bracket
screws (Right side)
Electrical
Receptacle
for AM
Doc. 065447-02 12/14
B-28).