Log Functions - Otto Bock B-Series Service Manual

Table of Contents

Advertisement

Parameter
Setting range
Minimum turn
1% increments
deceleration
of parameter
"Turn decelera-
tion value"
6.3.3

Log functions

The following functions are logged by the VR2 control and can be read with the corresponding hardware:
Function
Memory functions
Read timer
Clear timer
Read system log
Erase system log
B-Series
Description
Setting of the minimum turn deceleration of the wheelchair.
This percentage turn deceleration value appears if the speed has been set
to 1.
Programming example:
Turn Deceleration = 80 and Minimum Turn Deceleration = 25%
Turn Deceleration at step 1 = 25% of 80 = 20
Speed settings 2, 3 and 4 will interpolate linearly between 20 and 80
Turn deceleration at step 2 = 35
Turn deceleration at step 3 = 50
Turn deceleration at step 4 = 65
Description
VR2 is equipped with a timer and troubleshooting report function. They can be read and
deleted with the help of the DTT or PC programming device.
VR2 is equipped with a timer that records the hours of operation of the wheelchair. The timer
starts running as soon as the joystick is deflected from the home position and stops running
when the joystick returns to the home position. The timer records the number of hours for
which the wheelchair is in use.
This function resets the VR2 control's timer. This function can only be used with the PCPb
and PCPc versions of the PC programming device.
VR2 is equipped with a troubleshooting report system which stores the number of occurren-
ces of the last 8 detected system problems. This allows you to view the contents.
With this function the troubleshooting report from the VR2 control can be deleted. This func-
tion can only be used with the PCPb and PCPc versions of the PC programming device.
VR2 wheelchair control
Ottobock | 83