8.4.3 Programming of the Light Module

The light module is pre-programmed. By programming, the light module can be adapted to the
respective lights (see Section .9).
Programmable
Presetting
function
Operating voltage
2 V
of lamps
Lamp power
21 watts
Fault detection of
On (the control detects defects
direction indicators
on the direction indicator; is
indicated by quick flashing of
the direction indicator)
Table 12 Programming the parameters on the light module

8.4.4 Log Functions

The following functions are logged by the VR2 control and can be read with corresponding
hardware:
function
Description
Memory Functions The VR2 has a timer and a diagnostic log. These can be read and cleared us-
ing the PP1a programming device or the PC programming device.
Read Timer
The VR2 has a timer which records how long the wheelchair is in use. The timer
runs whenever the joystick is moved away from the centre position, and stops
when the joystick is returned. The timer records the number of hours the wheel-
chair has been in use.
Clear Timer
This function resets the VR2 control's timer. This function can only be made use
of with the PCPb and PCPc versions of the PC programming device.
Read System Log
The VR2 has a diagnostic log facility which stores the number of occurrences of
the last eight detected system problems. This allows you to view the contents.
Erase System Log This function clears the VR2 control's diagnostic log. This function can only be
made use of with the PCPb and PCPc versions of the PC programming device.
Table 13 Log functions
Possible adaptation
12 V (the lamps must correspond to the value
or must be replaced)
 watts; 10 watts
Off (no fault detection; use function only if one
direction indicator is used on each side)
7