Tips And Assistance; Questions And Answers - TESTO 330 Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 330:
Table of Contents

Advertisement

8

Tips and assistance

Pos: 88 /TD/Überschriften/8.1 Fragen und Antworten @ 0\mod_1177402017078_79.doc @ 1093 @ 2
8.1.

Questions and answers

Pos: 89 /TD/Tipps und Hilfe/Fragen und Antworten/testo 330 Fragen und Anworten @ 6\mod_1279266547844_79.doc @ 66975 @
Question
Rechargeable battery low
Measuring instrument
switches automatically off
or cannot be switch on
Displayed battery capacity
seems to be faulty
Error message: Pump
flow too high
Error message: Sensor
protection active
Error message: Printing
not possible
Possible causes / solution
> Switch to mains operation.
Batteries / rechargeable batteries
empty.
> Charge rechargeable batteries or
switch to mains operation.
Rechargeable battery was repeatedly
not fully discharged/charged.
> Discharge rechargeable battery
(until measuring instrument
switches off automatically) and
then charge fully.
Gas outlet blocked
> Make sure that the gas outlet is
clear.
The shutdown limit of the CO sensor
was exceeded.
> Remove the probe from the stack.
With printer 0554 0543: Bluetooth
interface not activated.
Wrong printer activated.
Printer switched off.
Printer outside wireless
transmission range.
> Activate Bluetooth interface, see
®
Bluetooth
, page 37.
> Activate printer used, see Printer,
page 36.
Switch on printer.
> Move printer within wireless
transmission range.
8 Tips and assistance
65

Advertisement

Table of Contents
loading