Sensor Protection; Recalibration / Adjustment - TESTO 330 Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 330:
Table of Contents

Advertisement

6 Using the product
6.1.3.3.

Sensor protection

Protection limits can be set to protect the sensors against overload.
Sensor protection switch-off is available for the following sensors:
CO, NO.
The sensor protection is activated when the threshold is exceeded.
testo 330-1 LL: Switch-off.
testo 330-2 LL: Dilution, if exceeded again: Switch-off.
To deactivate sensor protection the thresholds must be set to 0
ppm.
Calling up the function:
>
[
Setting sensor protection thresholds:
1. Select parameter: [Edit].
2. Set parameter → [OK].
3. Save changes: [Finished].
6.1.3.4.

Recalibration / adjustment

CO and NO sensors can be recalibrated and adjusted.
For recalibration / adjustment Testo recommends the use of the
calibration adapter 0554 1205.
Calling up the function:
>
[
Possibly:
> Enter the password:
[OK].
-
Gas zeroing (30 s).
40
]
Sensor Settings
If obviously unrealistic readings are displayed, the sensors
should be checked (calibrated) and, if required, adjusted.
Have the recalibration / adjustment carried out by a
qualified service centre approved by Testo.
Adjustments made with low gas concentrations can lead to
accuracy deviations in the upper measuring ranges.
]
Sensor Settings
[Enter]
Sensor Protection
Recalibration
→ Enter password →
→ [OK].
→ [OK].
[Next]