11 Tips and assistance
11.1 Questions and answers
Question
Possible cause
Instrument battery is
is displayed
almost spent
(top right on the
display)
•
Auto Off function is
Instrument
switches itself off
switched on
•
Remaining battery
capacity is insufficient
Display responds
Ambient temperature is
sluggishly
very low
Display: -----
Sensor error
Display: OOOOO
Permissible measuring
range has been exceeded
Display: UUUUU
Permissible measuring
range has been undershot
Display: BT Fail
Bluetooth connection
could not be established
Display: Print Fail
Printout could not be
performed successfully
Display: Probe Fail Probe damage
34
www.
Value
Instrument: 171 x 60 x 28 mm
Probe: 180 x Ø12 mm
Cable: 1.5 m
268 g
Possible solution
Replace instrument battery
•
Switch off Auto Off
function
•
Change the battery.
Increase ambient
temperature
Please contact your dealer
or Customer Service.
Keep within the permissible
measuring range
Keep within the permissible
measuring range
•
Check Bluetooth
connections.
•
Restart measuring
instrument, restart testo
Smart App.
•
Check Bluetooth
connections, switch off
and then on again if
necessary.
•
Switch printer off and then
on again.
Please contact your dealer
or Customer Service.
Testo-Direct
®
®