Assembly Replacement Sequence - Agilent Technologies E8362B Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Replaceable Parts

Assembly Replacement Sequence

Assembly Replacement Sequence
The following steps show the sequence that you should follow to replace an assembly in the
network analyzer.
Step 1. Identify the faulty group. Begin with
with the appropriate troubleshooting chapter that identifies the faulty assembly.
Step 2. Order a replacement assembly. Refer to this chapter.
Step 3. Replace the faulty assembly and determine what adjustments are necessary.
Refer to
Chapter 7, "Repair and Replacement Procedures."
Step 4. Perform the necessary adjustments. Refer to
Step 5. Perform the necessary performance tests. Refer to
Adjustments."
6-4
PNA Series Microwave Network Analyzers
Chapter 4, "Troubleshooting."
Chapter 3, "Tests and Adjustments."
Chapter 3, "Tests and
Service Guide E8364-90026
E8362B, E8363B, E8364B
Follow up

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363bE8364b

Table of Contents