System Recovery Procedure - Agilent Technologies E8362B Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Operating System Recovery

System Recovery Procedure

System Recovery Procedure
1. Click
,
Start
Shutdown
Normally, you can boot from the D: drive (recovery partition) which will allow
NOTE
you to erase and restore the C: drive. This is an available selection during the
normal boot-up process. If this boot-up selection process is not functioning
due to a damaged operating system, proceed to
Not Boot From the Hard Disk Drive" on page
2. After the Agilent logo displays on the screen, you will have several seconds to select an
alternative operating system from the default of Windows 2000.
a. Select the
Recovery
are booting from the recovery partition.
3. The analyzer will display a warning about proceeding and give you two chances to
abort. After the second warning, if not aborted, the restoration process begins
immediately.
Restoration takes from 30 to 60 minutes. There is no user interaction required. Near the
end of the process, the network analyzer will reboot several times. The network analyzer
application may start before the process is complete. Do not attempt to use the analyzer
until the network analyzer application is running and all activity has been stopped for at
least one minute.
Once the operating system is completely recovered, you can restore any files that you
saved. The mxcalfile_ files that were stored on the D:/Calfiles directory will be restored
automatically. Other files must be moved or copied manually.
NOTE
If the network analyzer options are lost, rename the C:/Program
Files/Agilent/Network Analyzer/gen.lic file to gen.bak, and then restart the
network analyzer application. This should regenerate the options
automatically and recreate the gen.lic file.
After recovery, you may notice that, upon power up, the system pauses for 30 seconds to
allow selection of the operating system. This time delay can be decreased to 5 seconds. To
do this:
1. Click
, point to
Start
2. Double-click
System
3. In the
System Properties
.
Recovery
4. Modify
Display list of operating systems for
value. The factory recommended setting is 5 seconds.
5. Click
in the
OK
Startup and Recovery
Close the
Control Panel
D-4
, and then
.
Restart
partition. A brief "splash" screen will appear indicating that you
, and then click
Settings
.
window, click on the
window. Click
window.
PNA Series Microwave Network Analyzers
"If the Operating System Does
D-5.
.
Control Panel
tab, and then click
Advanced
xx
where xx is the desired time
seconds.
in the
OK
System Properties
Service Guide E8364-90026
E8362B, E8363B, E8364B
Startup and
window.