Repair and Replacement Procedures

Removing and Replacing the Midweb and the B1 Fan

Removing and Replacing the Midweb and the B1 Fan
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• Pozidriv screw driver
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
1. Disconnect the power cord.
2. Remove the front panel assembly. Refer to
Assembly" on page
3. Remove the outer and inner covers. Refer to
4. On the bottom side of the analyzer, remove the A27–A30 first converters (mixers). This
is necessary to gain access to one of the midweb mounting screws. Refer to
and Replacing the A27, A28, A29, and A30 First Converters (Mixers) and Mixer Bias
Board" on page
7-50.
5. Remove the A13 (Option 080) frequency offset receiver assembly, if present. Refer to
"Removing and Replacing the A13 Frequency Offset Receiver Assembly" on page
6. Remove the A12 source 20. Refer to
page
7-22.
7. Remove the A11 phase lock board. Refer to
Lock Board" on page
8. Remove the A4 power supply assembly. Refer to
Supply Assembly" on page
9. Remove the card cage boards (A6, A8, and A10). Remove A9 (Option 080), if present.
Refer to
"Removing and Replacing the A6, A8, A9, and A10 Boards" on page
NOTE
This procedure has two parts: removing the midweb and removing the fan
from the midweb. You must remove the midweb before removing the fan.
7-72
7-10.
"Removing and Replacing the A12 Source 20" on
7-20.
7-16.
PNA Series Microwave Network Analyzers
"Removing and Replacing the Front Panel
"Removing the Covers" on page
"Removing and Replacing the A11 Phase
"Removing and Replacing the A4 Power
Service Guide E8364-90026
E8362B, E8363B, E8364B
7-6.
"Removing
7-24.
7-18.