Information In This Appendix; Appendix A At-A-Glance; Appendix D At-A-Glance - Agilent Technologies E8362B Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Error Terms

Information in This Appendix

Information in This Appendix

Your PNA series network analyzer calculates error terms by measuring well-defined
calibration devices over the frequency range of interest and comparing the measured data
with an ideal model for the devices. The differences of the compared data represent
systematic (repeatable) errors of the analyzer system. The analyzer calculates and stores
these differences as error terms. These error terms are then removed from subsequent
measurements when error correction is on.
NOTE
For details on the various levels and theory of error-correction, refer to the
embedded help in the analyzer.

Appendix A at-a-Glance

Section Title
Using Error Terms as a
Diagnostic Tool
Performing Measurement
Calibration
Using Flowgraphs to Identify
Error Terms
Accessing Error Terms
Error Term Data
A-2
Summary of Content
How to use error terms as a preventive
maintenance and troubleshooting tool.
Procedure for measurement calibration.
Measurement calibration must be performed
before you can use error terms as a tool.
Flowgraphs of the error terms for 1 port and
2 port calibrations.
How to access error terms either manually or
programmatically using the analyzer front
panel or SCPI and COM/DCOM commands.
Typical values of the error terms of your
analyzer. Included are:
• a table of typical values
• typical response plots
• typical assembly failures