Water connection
Pressurized DI
water connection
(30 - 1,000 kPa /
4.4 - 145 psi)
Non-pressurized
(gravity feed) DI
water connection
1.3 - 8.7 psi (8.5 -
60 kPa)
156
This machine can be connected to pressurized DI water with a
pressure between 30 - 1,000 kPa (4.4 - 145 psi). If the water pressure
is below 200 kPa (29 psi) the water intake duration will be
automatically increased.
The pressure tested hose for DI water, coded green, has a ¾ inch
female connection with garden hose thread and is connected to the
onsite DI water tap.
 If the machine is not going to be connected to DI water, the DI
water connection has to be deactivated by a Miele Service
technician. The inlet hose remains in position at the back of the
machine.
If DI water is supplied from a DI water reservoir with gravity feed or
low pressure, an optional DI water booster pump supplied by Miele is
required.
The pressure for gravity feed or low pressure ranges between 1.3 -
8.7 psi (8.5 - 60 kPa). The external DI water booster requires an
external power supply with 120V. For further details refer to install
diagram. The installation must only be carried out by Miele Service.