Dispensing / Dispensing systems

 Caution when handling process chemicals.
For all process chemicals, the process chemical manufacturer's safety instructions as
given on their safety data sheets must be observed.
Problem
Refill DOS
Prog. start not possible.
Prime dispenser pump DOS
Dispensing system DOS
priming
Priming DOS canceled.
Priming must be repeated
Possible cause and remedy
During a program sequence a low level of liquid process
chemical in a container has been identified.
Replace the empty container with a full one.
A program cannot be started because ...
... there is air in the dispensing system.
... the dispensing system has been sucked completely dry.
Check the fill level in the supply container. Replace the
empty container with a full one if needed.
Vent the dispensing system.
This is not a fault.
The dispensing system will now be automatically vented.
Wait until the venting process is finished.
Venting of the dispensing system was cancelled because
an insufficient flow rate has been identified. A dispensing
hose may be kinked or the suction lance blocked.
Check the dispensing hose for kinks and leaks. Position
it so that it cannot become kinked.
Check the suction opening of the suction lance for
blockages and remove these as necessary.
Start the venting process again.
Contact Miele Service if there are leaks in the dispensing
hose or a fault with the suction lance.
Problem solving guide
131