Problem
Spray pressure exceeds
tolerance
Spray pressure fluctuating too
much
Possible cause and remedy
The spray pressure has deviated by more than 20% from
the mean value.
Reasons for fluctuations in wash pressure are e.g.:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
The program was not interrupted, however the wash load
must be reprocessed again.
The program was cancelled because the spray pressure
has deviated by more than 40% from the mean value.
Reasons for fluctuations in spray pressure are e.g.:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
Reprocess the load again.