Miele PG 8593 Operating Instructions Manual page 101

Laboratory glassware washer
Table of Contents

Advertisement

Activating
conductivity
measuring
Set limit value
Conductivity measuring is activated individually for each program.
Conductivity is measured in the final rinse phase.
   ...
Conductivity limit
         
Conductivity limit
Water intake
Water drainage
Water intake
The electrical conductivity of the water before the beginning of the
final rinse phase is measured.
Water drainage
The electrical conductivity of the water at the end of the final rinse
phase is measured.
Select an option using the  and  arrow buttons.
Press OK to save the setting.
Following selection a maximum limit value for electrical conductivity
must be set.
The limit values for
individually.
   ...
Conductivity limit
         
Water intake
             
Set
                 
Water intake
µS/cm
--
(5 - 200 µS/cm )
Adjustment is in increments of 1. The possible range is shown in the
bottom line.
Use the arrow buttons  (higher) and  (lower) to set the limit
value.
Touch OK to save the limit value.
Water intake
Water drainage
and
Water drainage
or
Program settings
are set
Water drainage
--
µS/cm
(5 - 200 µS/cm )
101