Miele PG 8593 Operating Instructions Manual page 115

Laboratory glassware washer
Table of Contents

Advertisement

Problems with
data transmission
Communication module for external archiving
A module slot is integrated into the back of the machine for a Miele
communication module for permanent archiving of cycle reports. The
module enables the installation of an Ethernet interface for
documentation via documentation software or a RS-232 interface to
connect a report printer.
Please contact Miele for further information on software for process
documentation and suitable printers.
Only use terminal devices (PC, printers etc.) which comply with
CAN/CSA-C22.2 No. 60950-1 or UL 60950-1.
Communication modules are available from Miele and can be fitted
retrospectively at any time. The modules come with their own
Installation instructions.
The interface must be configured by suitably qualified and competent
personnel. Please refer to the information in "Further settings /
Interface".
Process documentation using external software (option)
For digital archiving the process data is transmitted to external
process documentation software via an Ethernet interface.
Transmission can optionally occur continuously during the process or
as a single data packet at the end of the process. The settings for this
are modified by Miele Service.
Information on wash pressure, A
in the wash cabinet can be archived graphically if required.
Installation of an Ethernet interface requires the retrospective fitting of
an XKM 3000 L Med communication module.
For connection to a WLAN network the module can be connected via
a cable to an existing wireless access point.
If there is a network problem during a running process, e.g. due to a
loose cable, a relevant fault message is displayed.
Network down
The process running will be continued without interruption and the
process data will be saved in the meantime in the internal memory.
In the event of network or report software problems contact your
system or network administrator.
Process documentation
value, conductivity and temperature