Miele PG 8593 Operating Instructions Manual page 132

Laboratory glassware washer
Table of Contents

Advertisement

Problem solving guide
Problem
Check container/lance DOS
Highly viscous (thick) process chemicals can affect the dispenser monitoring and lead to
inaccurate data. In this instance please contact Miele Professional Service for advice.
132
Possible cause and remedy
Little or no flow has been identified.
Check the fill level in the supply container. Replace the
empty container with a full one if needed.
Check the suction aperture of the suction lance for
deposits.
Vent the dispensing system.
The dispensing hose is kinked.
Remove any kinks from the dispensing hose. Position it
so that it cannot become kinked.
Check the dispensing hose for leaks.
Vent the dispensing system.
Contact Miele Service if there are leaks in the dispensing
hose or a fault with the suction lance.