2.1 Site Requirements

Required Space
Before beginning the installation procedure, note the following autosampler site and utility
requirements
(Figure
Note:
The autosampler site requires room for a waste bottle and a needle wash solvent
reservoir.
• Bench space – Allow 21 x 23.5 inches (53 x 72 cm) for the autosampler. This space
includes ventilation sides 2.5 inches (6.3 cm) and back 3.5 inches (8.75 cm).
Note:
Ensure the bench top must be level to ensure proper drainage of the solvent
spill lines.
• Bench space with heater/cooler installed – Allow an additional three inches on the
back and on each side next to the air vents.
Note:
To avoid heater/cooler performance degradation, do not place the rear of the
unit in an enclosed space, under a shelf, or near a heat source.
• Column location – Ensure the red fluid line reaches the column. Lengthening this
line may result in peak broadening.
• Waste and solvent container location – Place the needle wash reservoir at bench
level to prevent siphoning.