Refilling The Alcohol Cartridge With Alcohol - Kanomax 3300 Operation Manual

Portable aerosol mobility spectrometer
Table of Contents

Advertisement

2. Getting Started

2.2.2 Refilling the Alcohol Cartridge with Alcohol

Do not leave the alcohol cartridge inlet open.
Caution
Failure to observe the above may cause contamination of the optical
system or a malfunction.
1. Turn the instrument off.
2. Open the alcohol bottle by turning the storage cap (or the
alcohol cartridge) about 45° counterclockwise.
Stand the storage cap (or the alcohol cartridge) straight up in
a clean place.
3. Pour isopropyl alcohol in the alcohol bottle up to the marked
level. Be careful not to tip the bottle and spill the alcohol.
4. Insert the alcohol cartridge into the alcohol bottle, and turn it
about 45° clockwise until it is firmly locked.
5. After the alcohol cartridge is set, the felt in the cartridge will be soaked in alcohol.
You can use the instrument after a few minutes of soaking the felt in the alcohol.
Alcohol Fill Level
6