7. Troubleshooting
The charging level of
the corona current
does not reach its
maximum, and a
measurement
cannot be started.
The message
"Please do
maintenance on the
charger." appears on
the screen.
Symptom
[FLOWERR]
The inlet (cyclone) is clogged.
message
The alcohol cartridge or storage
cap is not firmly inserted.
[MEMERR]
The internal flash memory runs
message
out of memory.
[TILTERR]
The instrument is tilted.
message
[COMERR]
An internal cable is
message
disconnected.
[LDPDERR]
A problem has occurred with the
message
optical system.
[PUMPERR]
A problem has occurred with the
message
pump.
[PELTIER]
A problem has occurred with the
message
peltier device.
Deposits have been built up
on the electrodes in the
charger.
Possible Cause
On the Menu screen, tap the
Maintenance button to refresh the
charger. If several refreshes do
not work, contact your distributor
or the KANOMAX service center
(see the last page of this manual).
Troubleshooting
Check the inlet (cyclone) for clogs
and clean it if necessary.
(Refer to 5.2 Inlet (Cyclone).
Insert the cartridge or storage case
and turn it until it can't be turned
any further.
Select [MENU][File].
Delete the data file or move the data
file to a USB flash drive.
The pump stops working if the
instrument is tilted.
Make sure the instrument is level.
Contact your distributor or the
KANOMAX service center (see the
last page of this manual).
Contact your distributor or the
KANOMAX service center (see the
last page of this manual).
Contact your distributor or the
KANOMAX service center (see the
last page of this manual).
Contact your distributor or the
KANOMAX service center (see the
last page of this manual).