Appendix
The data file format, which consists of the header part (e.g. measurement settings) and the
data part, is as follows:
[Header part]
;WaitTime,20
;DelayTime,0000
;MeasTime,0010
;DeadTime,0005
;LoopWait,0020
;MeasTimes,9999
;Corona,-5.00,05.00
;WSP14191
;WSP2-533
;WSP31542
;DMA00,0004,0011.55,7272.73
;DMA01,0007,0015.40,1538.46
;DMA02,0013,0020.54,0496.89
;DMA13,2951,0486.97,0073.66
[Data part]
Time,Diameter,Inverted,RAW,DMA,Corona,Corona,DMA(meas),Error,Qa,Inlet,Qsh,Temp1,Temp2
[Data sample]
20:08:08, 11.6, 7273, 1, 4, -3.3, 5.1, -5.5, 00000000, -532, 4192, 1474, 24.0, 0.4
20:08:18, 15.4, 4615, 3, 7, -3.4, 5.1, -9.5, 00000000, -534, 4203, 1601, 24.2, 0.4
Wait time before starting a measurement (second)
(Not in use)
Measurement time per channel (second)
(Not in use)
Wait time from a completion of a scan to starting a next scan (second)
Number of scans
Setting range of corona current (not in use)
WSP1-3: Setting value of the flow rate (not in use)
DMA0-13 (High Resolution Mode)
DMA0-26 (Wide Scan Mode)
Temperature inside the main unit 2
Temperature inside the main unit 1
Differential Pressure (DP4) (200cc)
Differential Pressure (DP3) (700cc)
Differential Pressure (DP1) (50cc)
Setting values of voltage and particle size for each channel of DMA
Error Code
DMA Voltage (Measurement: V)
Corona Current (µA)
DMA Voltage (Parameter: V)
Concentration (RAW Data)
Concentration (Inverted)
Particle Size (nm)
Time (hh:mm:ss)
38