Kanomax 3521 Operation Manual

Kanomax 3521 Operation Manual

Piezobalance dust monitor
Hide thumbs Also See for 3521:

Advertisement

Quick Links

Piezobalance Dust Monitor
MODEL 3521
Operation Manual
PIEZOBALANCE DUST
MONITOR MODEL 3521
Read this manual carefully and understand the warnings
before operating the instrument.
Keep this manual handy for future reference.
03002
11 .10

Advertisement

Table of Contents
loading

Summary of Contents for Kanomax 3521

  • Page 1 Piezobalance Dust Monitor MODEL 3521 Operation Manual PIEZOBALANCE DUST MONITOR MODEL 3521 Read this manual carefully and understand the warnings before operating the instrument. Keep this manual handy for future reference. 03002 11 .10...
  • Page 2: Important Safety Information

    Important Safety Information Types and definitions of warning signs used in this operation manual are shown below. Danger: To prevent serious injury or death. Items under this heading show measures to prevent serious injury or death, which may result if the instructions in this manual are not observed and the instrument is operated inappropriately.
  • Page 3 Caution * Do not drop or strike the instrument. …… Dropping or hitting the instrument may Prohibition cause damage or a malfunction. * Never disassemble, modify or repair the instrument . …… Failure to observe the above may cause a short circuit Do not modify / disassemble or a malfunction.
  • Page 4: Table Of Contents

    Table of Contents 1. Part Names and Functions ..................1 1.1 Part Names........................1 1.2 LCD Screen ........................2 1.3 Operation Keys ........................ 2 2. Battery Indicator......................3 3. Battery Charge ......................4 3.1 AC/DC Adapter........................ 4 3.2 Battery Charging Display ....................4 4.
  • Page 5 10. Troubleshooting....................23 11. Warranty and After-sales Service ...............24 12. Contact Information ....................26...
  • Page 6: Part Names And Functions

    1. Part Names and Functions 1.1 Part Names Impactor (unit: mm (inches)) Needle Display screen PIEZOBALANCE DUST MONITOR MODEL3521 Operation keys 180 mm (7.1”) Display screen Communication port contrast adjustment Cleaning port Power switch Rechargeable battery pack AC adapter socket Cleaning device Battery pack release...
  • Page 7: Lcd Screen

    1.2 LCD Screen a. Date & Time: Indicates current date & time. b. Sampling Time: Sampling time can be quick set to either 24s, 120s or custom set from 1-60 min. c. Data Record: Indicates the current number of data records (Max.500). When a measurement is complete, “STORE”...
  • Page 8: Battery Indicator

    2. Battery Indicator Remaining battery life is displayed when the instrument is powered on. The battery indicator changes as shown below: Automatic Requires Battery indicator Full Half power off recharge flashes The battery indicator flashes when the battery voltage drops below 9.3V. The AC/DC adapter must be connected immediately to continue measuring.
  • Page 9: Battery Charge

    3. Battery Charge 3.1 AC/DC Adapter For prolonged measurements, the instrument should be powered using the supplied AC/DC adapter (when AC100-240V power is available). Insert the adapter jack into the adapter socket on the instrument first, and then to a power outlet. Adapter jack Adapter Adapter...
  • Page 10: Parts Installation And Measurement Preparation

    4. Parts Installation and Measurement Preparation 4.1 Removal and Installation of the Rechargeable Battery Pack 1) Press the rechargeable battery pack release, and slide the battery pack along the guiding groove away from the release as shown below. Battery pack release Rechargeable battery pack Guiding groove 2) Align the tabs on the rechargeable battery pack with the slots in the guiding groove and lift the...
  • Page 11: Accessing And Storing The Cleaning Device

    4.2 Accessing and Storing the Cleaning Device 1) To access the cleaning device, hold the cleaning device while sliding the lock away from the cleaning device as shown below. Cleaning device Cleaning device lock Rechargeable battery pack 2) After the lock is released, remove the cleaning device by lifting the front end. Front end of the cleaning device 3) To return the cleaning device to storage, place the rear end of the cleaning device onto the...
  • Page 12: Confirming The Sensor Condition

    4.3 Confirming the Sensor Condition When the instrument is powered on, the screen below will display a flashing “WAIT” message in the operating status field. After approx. 10 seconds the sensor will stabilize and the “WAIT” message will stop flashing. The measurable range of the sensor can be confirmed by the bar indicator and the frequency displayed on the screen.
  • Page 13: Cleaning The Sponge

    4.4 Cleaning the Sponge Preparation Before Cleaning 1) The sponge on the cleaning device must be moistened before use. If the sponge is dirty or stained, put the cleaning device into a vessel filled with clean water and a few drops of neutral detergent, and let it soak for a few minutes (this is not necessary for a new product).
  • Page 14: Cleaning The Sensor

    4.5 Cleaning the Sensor First, confirm that the high voltage is turned off. Then align the front end of the cleaning device with the cleaning port, and slowly insert the cleaning device with the sponge side facing down. The cleaning device must be inserted up to the scale line indicated on the back side of the cleaning device.
  • Page 15: Adjusting The Impactor Position

    4.6 Adjusting the Impactor Position The impactor may be set at three angles: 0°, 45° and 90°. Turn the impactor to the desired angle. (Note: The instrument has built in stops for each angle.) Rotation angle: 45° Rotation angle: 90° Rotation angle: 0°...
  • Page 16: Measurement Method

    5. Measurement Method 5.1 High Voltage Supply When the instrument is turned on, the flashing “WAIT” message will be displayed. After a brief delay, the “WAIT” message will stop flashing, and high voltage can be supplied. Press [MEAS/CLN] to supply high voltage. The pointer at the left end of the high-voltage indicator frame will move toward the right.
  • Page 17: Measurement Start

    5.2 Measurement Start To start a measurement, press [START] after the operating status has changed to “READY”. The “READY” message will change to an elapsed time indicator. During the measurement, the bar indicator will extend as time elapses and the concentration reading will blink. The measurement will stop automatically when the set time has elapsed.
  • Page 18: Measurement End

    5.3 Measurement End When a measurement is complete, the concentration reading will stop flashing. The displayed value is the measurement result for the specified sampling time. The operating status display will change from the elapsed time indicator to the “READY” message, and the number of data records will change to a flashing “STORE”...
  • Page 19: Settings Menu

    6. Settings Menu 6.1 Main Menu The following menu will be displayed by pressing [MENU] (when high voltage is not being supplied to the instrument). Use △ and ▽ to select an item. To enter the sub-menu of each item, press [SET]. To return to the measurement screen, select “1.
  • Page 20: Data Output Menu

    6.3 Data Output Menu To select and view stored data Move the cursor to item “3. DATA OUTPUT” on the main menu by using △ and ▽ then press [SET]. The screen shown below will be displayed. If there are no stored data records, the message “There is no record. Please set MENU key” will be displayed.
  • Page 21: Communications Menu

    it is recommended that you set the time for approx. 30-60 seconds ahead of the actual time, and then confirm the setting by pressing [SET] on “3.SAVE INFO” when the set time matches the actual time. 6.5 Communications Menu To set the communications baud rate Move the cursor to item “7.COMMUNICATION”...
  • Page 22: Data Clear Menu

    6.6 Data Clear Menu To delete stored measurement data Move the cursor to item “5.DATA CLEAR” on the main menu by using △ and ▽ then press [SET]. The screen shown below will be displayed. If there are no data records, a message “There is no record, Please set MENU key”...
  • Page 23: Printing

    7. Printing 7.1 Connecting the Printer Connect the 3521 to the optional printer using the Communication optional printer cable. (Refer to the operation 3521 cable manual of the printer for details.) Set the communication baud rate to 9600 bps. (The factory default setting is 9600 bps, so it is not necessary to set it unless it has been changed by the user.)
  • Page 24: Regular Maintenance And Impactor Nozzle Replacement

    8. Regular Maintenance and Impactor Nozzle Replacement Regular maintenance is required to ensure the long-term accuracy and performance of the instrument. In order to maintain the initial accuracy, it is recommended that the sensor and impactor plate be cleaned before and after use. If the instrument is used for a long time, or used in a high concentration environment, periodical cleaning of the needle is recommended.
  • Page 25: Removing And Cleaning The Needle

    Do not dip the upper portion of the needle in the detergent. Before re-inserting the needle into the instrument, ensure it is completely dry or damage to the instrument may occur. Needle PIEZOBALANCE DUST MONITOR MODEL 3521 Cleaning solution made up of 1/2 of the provided detergent and 1/2 purified water...
  • Page 26: Installing The Needle

    8.3 Installing the Needle To reinstall the needle, look into the inlet from above and align the high-voltage terminal on the side of the needle with the groove inside the inlet, then slowly twist in the needle. When the needle is fully inserted, lock it in place by turning the knob to the “S” mark until it cannot be turned any further.
  • Page 27: Main Specifications

    9. Main Specifications Product Name Piezobalance Dust Monitor Model 3521 Mass concentration of suspended particulate matter in indoor or Measuring Object outdoor air. Operating Temperature Range -10°C to 40°C (14°F to 104°F) Storage Temperature Range -30°C to 60°C (-22°F to 140°F) Measuring Range 0.01 to 5mg/m...
  • Page 28: Troubleshooting

    10. Troubleshooting Please review the following troubleshooting tips before requesting a repair. Symptom Possible Cause Solution Turn OFF the power and charge the The display does not appear Battery completely discharged. battery. when the power is turned Contact failure of the battery Remove and reinstall the battery pack.
  • Page 29: Kanomax Limited Warranty

    11. Warranty and After-sales Service KANOMAX Limited Warranty The limited warranty set below is given by KANOMAX with respect to the KANOMAX brand Piezobalance Dust Monitor (Model 3521) and its attachment parts including accessories (hereafter referred to as “PRODUCT”) that you have purchased. PRODUCT you have purchased shall be the only one that the limited warranty stated herein applies to.
  • Page 30 KANOMAX. THIS LIMITED WARRANTY SHALL NOT EXTEND TO ANYONE OTHER THAN THE ORIGINAL PURCHASER OF THE PRODUCT, OR THE PERSON FOR WHOM IT WAS PURCHASED AS A GIFT, AND STATES THE PURCHASER'S EXCLUSIVE REMEDY. After-sales Service If the PRODUCT is malfunctioning, please check with “Troubleshooting” to find possible causes first.
  • Page 31: Contact Information

    12. Contact Information JAPAN & ASIA KANOMAX JAPAN, INC. 2-1 Shimizu Suita City, Osaka 565-0805, Japan TEL: 81-6-6877-0183 FAX: 81-6-6879-5570 http://www.kanomax.co.jp/ URL: sales@kanomax.co.jp E-Mail: USA & EUROPE KANOMAX USA, INC. PO Box 372, 219 Route 206, Andover, NJ 07821 U.S.A.

Table of Contents