3. Operation Procedures
Wait Time
Meas Time
No. of Scans
Y Scale
Graph Y Max
Graph Y Min
Data Folder
Para Folder
Sheath Flow
[Measurement Time]
If the settings are as follows, the whole measurement time can be calculated as shown
below:
(example) Wait Time: 20 seconds
Meas Time: 10 seconds/ch
No. of Scans: 10 scans
Sheath Flow: Wide Range(14ch)
Loop Wait: 20 seconds
*Loop Wait: wait time set for each scan (unchangeable)
Measurement time
= 15 seconds+(20 seconds+10 seconds/chx14ch)x10=1,570 seconds
Sets a wait time before starting a measurement. Settable range is
from 0 to 600 seconds.
Sets a measurement time per channel (ch). Settable range is
from 2 to 600 seconds/ch.
Sets the number of scans. Settable range is from 1 to 1,000 scans.
Select Auto to change the Y axis of the graph automatically
according to the measurement results.
Select Manual to fix the Y axis of the graph regardless of the
measurement results. Tap Auto or Manual to select either of
them.
Sets the maximum value of the Y axis of the graph. This parameter
is valid when the Y Scale is set to Manual. The settable range is
from 1 to 30,000,000 pcs/cc.
Sets the minimum value of the Y axis of the graph. This parameter
is valid when the Y Scale is set to Manual. The settable range is
from 0 to 29,999,999 pcs/cc.
Assign a folder to store the measurement results. When a USB
flash drive is installed, this Data Folder will be created on the USB
flash drive. The factory default folder is SCAN.
Assign a folder to record the parameters. When a USB flash drive
is installed, this Para Folder will be created on the USB flash drive.
The factory default folder is PAMS.
In the High Resolution mode, a range from 10nm to 433.7nm is
divided into 27 classifications to perform measurements with a high
accuracy.
In the Wide Range mode, a range from 14.5nm to
862.3nm is divided into 14 classifications to perform quick
Select the High Resolution or Wide Range by
measurements.
tapping either of them.