In The Event Of Memerr; In The Event The Charger Is Not Ready - Kanomax 3300 Operation Manual

Portable aerosol mobility spectrometer
Table of Contents

Advertisement

3. Operation Procedures
Selectable page range

3.3.3 In the Event of MEMERR

3.3.4 In the Event the Charger is Not Ready

Up to 100 data from the measurement
results can be displayed per page.
you set the number of measurements to
more than 100 in the [No. of Meas]
settings, the results will be displayed
over multiple pages.
When there are multiple graph pages,
tap the page selection to display the
page selection dialog box.
page number you wish to display.
The selectable page range is shown on
the title bar of the dialog.
Tap [Prev] button to return to the
SCAN-tab screen.
When the data storage space runs out,
[MEMERR] will be displayed.
In the event this error is displayed,
select [MENU]  [File] and delete some
stored data file(s) or move some data
file(s) to a USB flash drive.
After turning the power ON, if the
charger is not ready within 15 minutes,
the message shown to the left will
appear on the screen.
If this message appears, please refer to
3.4
MENU,
Maintenance
maintenance on the charger.
20