3. Operation Procedures
· Maintenance
Charger refresh allows the corona current to approach to its normal level:
Charger Current(N)= -3.0±1.0uA
Charger Current(P)= 3.0±1.0uA
Tap the [Start] button to refresh the charger. The [Start] button will be replaced with
the [Stop] button. Approximately ten (10) minute later, tap the [Stop] button to finish
the process.
The Maintenance window allows you to
refresh the charger.
Over time, deposits accumulate on the
electrodes in the charger, making the
corona current unstable and preventing
a measurement from getting started.
(Refer to 3.3.4 In the Event the
Charger is Not Ready.)
After turning the power ON, if the
charger is not ready within
15 minutes, the message "Please do
maintenance on the charger." will
appear on the screen. If this message
appears, you can perform maintenance
on the charger from the Maintenance
menu.