Maintenance Cycle - Kanomax 3300 Operation Manual

Portable aerosol mobility spectrometer
Table of Contents

Advertisement

5. Maintenance

5.3 Maintenance Cycle

To operate this instrument correctly, please follow the recommended maintenance
shown below. We highly recommend you refill the alcohol cartridge with alcohol every
time before using the instrument. Conduct other maintenance, as necessary depending
on the frequency of use and/or condition of the instrument.
Items
Alcohol Cartridge
Felt inside the Alcohol
Cartridge
Alcohol Cartridge
Insertion Opening
Zero Filter
Inlet (Cyclone)
Charger
Main Unit
Maintenance
Soak the cartridge in the
alcohol.
Replace the felt inside the
alcohol cartridge.
Wipe off any alcohol
accumulated inside the
cartridge opening.
Attach the zero filter to the
inlet (cyclone).
Clean the inlet (cyclone).
On the Menu screen, tap the
Maintenance button to
refresh the charger.
Return the instrument to
your distributor for cleaning
and calibration.
Frequency
Daily (or every time before using)
(Refer to 2.2 Refilling the
Alcohol Cartridge with
Alcohol)
When necessary
(Refer to 5.1 Alcohol Cartridge
and 7.Troubleshooting.)
Before using the instrument
(Refer to 2.2 Refilling the
Alcohol Cartridge with
Alcohol)
After using the instrument
Attach the zero filter and allow
about 5 minutes before turning
the power off.
(Refer to 3.1 Boot and
Shutdown)
When necessary
(Refer to 5.2 Inlet (Cyclone)
and 7. Troubleshooting)
When the message "Please do
maintenance on the charger."
appears on the screen.
When the charging level of the
corona current does not reach its
maximum.
(Refer to 3.4 MENU)
Once a year
31