Process Step: Wash - BIO RAD Bio-Plex Pro Instructions For Use Manual

Wash stations
Table of Contents

Advertisement

Defining New Programs

Process Step: Wash

This process step is used to aspirate solution out of the wells and simultaneously
dispense and aspirate the solution.
58
CX PX W ASH
-
+
EXIT OK
CROSSW. ASP.
NO
-
+
EXIT OK
POS: OVERFLOW
-
+
EXIT OK
VOLUME 300µl
-
+
EXIT OK
CHANNEL 1
-
+
EXIT OK
H-SPEED 10 mm/s
-
+
EXIT OK
FL.RATE 200 µl/s
-
+
EXIT OK
ASP. RATE
2
-
+
EXIT OK
TIME
1 s
-
+
EXIT OK
CX PY WASH
-
+
EXIT OK
Select and confirm wash step (W ASH)
CX = Cycle Number
PX = Program Step Number
Select Yes using – or + to perform aspiration of
wash crosswise or select No to perform normal
aspiration.
Select required dispensing position
(OVERFLOW, CUSTOM, MOVE OVER., MOVE
CUST. or BOTTOM).
Select the volume of solution to be used
(50 - 3000 µl in increments of 50 µl)
Select the channel to be used (1 - 4)
depending on instrument configuration.
Select required aspirating speed (1 - 20 mm/sec).
Select required flow rate (drip to 500µl/sec).
Select required aspirating rate (1 – 3).
Select required aspirating time (1 – 20 mm/s).
After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another wash step
or press – or + to select the next program step,
Program End or a new cycle.
Bio-Plex Pro and Bio-Plex Pro II Wash Stations