If the instrument will be switched off at the end of operation, the Rinse Night
procedure should be performed. This prevents the needles from becoming blocked.
Use the Rinse Night procedure to rinse the wash system as outlined above.
When the rinsing procedure is completed the following message is displayed:
The manifold remains in the prime tray until END is pressed. Press STOP to abort
the procedure (prime tray will not be aspirated).
Changing the Plate Carrier
See Page 16 for detailed instructions on selecting and replacing the plate carrier.
Bio-Plex Pro and Bio-Plex Pro II Wash Stations
RINSE NIGHT
STOP END
The most important cleaning procedure for this instrument is rinsing the
liquid system with distilled water at the end of each day.
Basic Operation and Maintenance
27