22.1 Technical data

Generator
X-ray tube
Focal spot size
Cone diameter
Max. symmetrical radiation field
Total filtration
Inherent filtration
Anode voltage
Anode current
Target material
Target angle
Exposure times
Reference current time product
Lowest current time product
Max. nominal anode voltage
Power input
Max. electrical output
Electrical output at 0.1 sec.
Max. loading energy
SID
(Source - Image receptor
Distance)
SSD (Source-Skin Distance)
Standard/Long
Long with rectangular collimator
Mains voltage
Apparent resistance
Mains frequency
42
Planmeca ProX
Constant potential, microprocessor controlled,
operating frequency 66 kHz
Toshiba D-041SB
0.4 mm according to IEC 60336
60 mm (2.36 in.)
Rectangular 36 x 45 mm (1.42 x 1.77 in.)
ø 60 mm at SSD 200 mm
ø 60 mm at SSD 300 mm
according to IEC 806
min. 2.5 mm Al equivalent at 70 kV
according to IEC 60522
1 mm Al equivalent at 70 kV, according to IEC
6052
2-8 mA: 60, 63, 66, 70 kV, ±2 kV
8, 7, 6, 5, 4, 3, 2 mA, ± (5% + 0,2 mA)
Tungsten
12.5°
0.01- 2 sec. ±(5% + 0.001 sec.), 24 steps
8 mAs at 70 kV, 8 mA, 1 sec.
0.02 mAs at 2 mA, 0.01 sec.
70 kV
1000 VA (220-240 V)
890 VA (100-115V)
560 W at 70 kV, 8 mA
560 W at 70 kV, 8 mA
1987 mAs/h at 70 kV
min. 200 mm (8 in.)
200 mm (8 in.)/ 300 mm (12 in.)
306 mm (12.04 in.)
100 V~/220-240 V~
0.3 ohms 100-115 V~ /
0.8 ohms 220-240 V~
50/60 Hz