16 PROGRAMMING EXPOSURE VALUES

16 PROGRAMMING EXPOSURE VALUES

16.1 Programming default exposure values

34
Planmeca ProX
The default exposure values can be programmed for both
the adult and child mode. The indicator light of the
selected projection will come on. The current exposure
values are shown on the time, kV and mA displays.
NOTE
Make sure that no exposure region is selected, i.e. no
preprogrammed setting indicator light is on.
NOTE
The exposure parameters - time, kV and mA - are
programmed according to the density value 0. The time
value will be automatically changed according to the
selected density value in the film-based imaging mode, in
the digital imaging mode and in the imaging plate mode
when you exit the programming mode.
Press and hold down the SELECT key (about 4 seconds)
until you have heard a signal tone to enter the
programming mode.
The imaging mode can be changed by pressing the
MODE key briefly.
The kV range can be modified in Service mode. For more
information see Planmeca ProX Technicl manual.
The ready light will start to flash. The time display will start
to flash and the default exposure values will appear on the
displays.
The exposure time value is changed with the parameter
adjustment keys. The exposure times are shown in
section 17 "EXPOSURE VALUE TABLES" on page 38.
Press the SELECT key briefly, the kV display will start to
flash and the exposure time value is stored in the
memory.
The kV value can now be changed with the parameter
adjustment keys.
Press the SELECT key briefly, the mA display will start to
flash and the kV value is stored in the memory.
The mA value can now be changed with the parameter
adjustment keys.