Page 1
Planmeca ProX ™ ™ for veterinary use user's manual 30028127...
Page 2
IEC 60364 - equipment is used according to the operating instructions. Planmeca pursues a policy of continual product development. Although every effort is made to produce up-to-date product documentation this publication should not be regarded as an infallible guide to current specifications. We reserve the right to make changes without prior notice.
This manual is valid for software revisions 4.00 or later. NOTE The Planmeca ProX intraoral X-ray unit is allowed to be used only under supervision of a dental/health care professional. The Planmeca ProX intraoral X-ray unit fulfils the requirements of Directives 93/42/EEC and 2011/65/EU (RoHS).
Planmeca ProSensor installation manual • Planmeca ProSensor technical manual These manuals are intended to be used in conjunction with the documentation for the Planmeca Romexis imaging software. The imaging software package contains the following manuals: • Planmeca Romexis user’s manual •...
To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth. Intermediate focal spot (Standard IEC 60417). Separate collection for electrical and electronic equipment according to Directive 2002/96/EC (WEEE). No pushing (Standard ISO 7010). User's manual Planmeca ProX...
Make sure that you are fully acquainted with the appropriate radiation protection measures and these instructions before using the unit. WARNING Failure to install the Planmeca ProX X-ray unit in an approved location may be dangerous to both patient and operator. WARNING No modification of this equipment is allowed.
Page 9
This X-ray unit may be dangerous to both patient and operator unless safe exposure values are used and correct operating procedures are observed. CAUTION The SIP/SOP shall not be used in Planmeca ProX, but only for connecting Planmeca ProSensor device. CAUTION Do not connect a multiple portable socket outlet (MPSO) or extension cord to the system.
CAUTION Do not connect any other equipment to the control panel’s terminal. 5.2 Mobile base For more information on transport position and transportation instructions, see section "Transportation (mobile base)" on page 33. Planmeca ProX User's manual...
Page 11
CAUTION The maximum allowed weight on the optional mobile tray is 3 kg (7 lbs). CAUTION The Planmeca ProX mobile X-ray unit has to be in transport position when on a surface inclined 5° or more. User's manual Planmeca ProX...
Tube head CPU software version on the time display. After the self-test is completed, the default exposure values (if set) appears on the displays. The default exposure values can be reprogrammed by the user, see section "Programming exposure values" on page 23. Planmeca ProX User's manual...
1 Attach/remove: Red points in line 2 In position: Red and black points in line User's manual Planmeca ProX...
Page 14
(2). The cone can be now rotated in its position ±90°. The rectangular cone can be removed when the red dots on the short cone and on the rectangular cone are in line. Planmeca ProX User's manual...
The selected kV value is shown on the kV display. There are four different values that can be selected: 60, 63, 66 and 70 kV. NOTE The kV range can be 60-70, 66-70, 60-68, 66-68 or 68 depending on the local requirements. User's manual Planmeca ProX...
The indicator light of the selected projection will come on. Pressing the key a third time will recall the default exposure values. The preprogrammed settings can be changed by the user, see section "Programming exposure values" on page 23. User's manual Planmeca ProX...
- to be changed. When the parameter value is flashing on the display, the parameter can be changed with the parameter adjustment keys. The up key increases the value and the down key decreases it. Planmeca ProX User's manual...
You can set the unit so that the Ready indicator light will only come on when Waiting for Planmeca Romexis is ready for the exposure, i.e. the message exposure appears on the computer screen. To change the settings of the unit contact your Planmeca technical support.
The indicator light of the selected projection will come on. 1 Incisors 2 Premolars and canines 3 Molars 4 Occlusal exposure 5 Endodontic exposure 6 Bite-wing exposure Planmeca ProX User's manual...
After adjusting the kV or mA value the unit will return automatically to the time adjustment mode after 5 seconds time. 9.1 Recommended exposure values for cats or smaller animals The following table shows the recommended exposure values for Planmeca ProSensor. NOTE These values are for digital sensor.
5 Bitewing (BW) Not used for animals (or use the same parameters as for molars). 9.2 Recommended exposure values for larger animals The following table shows the recommended exposure values for Planmeca ProSensor. NOTE These values are for digital sensor.
Page 23
5 Needle in root Exposure kV/mA: 60 kV, 8 mA canal Exposure time: 0.250 s Note, fine tune the exposure parameters depending on teeth. 5 Bitewing (BW) Not used for animals (or use the same parameters as for molars). User's manual Planmeca ProX...
Dogs and cats do not have an arched palate and that is why the maxillary teeth cannot be imaged with the paralleling technique. • The mandibular symphysis interferes the sensor placement parallel to the tooth roots of the mandibular canines and incisors as well as the rostral mandibular premolars. Planmeca ProX User's manual...
Page 25
10 Sensor positioning In the paralleling technique: • The sensor (2) is placed parallel to the object being radiographed, and the X-ray beam is directed perpendicular to both the sensor and the tooth/root (1). User's manual Planmeca ProX...
3. Press and hold the exposure key on the control panel for the duration of the exposure. Results The exposure warning light will come on. You will also hear the radiation warning tone during the exposure. Planmeca ProX User's manual...
1. Select the large or small animal mode of which the default exposure values is to be programmed. The indicator light of the selected mode will come on. The current exposure values are shown on the time, kV and mA displays. User's manual Planmeca ProX...
Page 28
The default kV value can now be changed with the parameter adjustment keys. Note, the available kV range can be modified in Service mode, for more Planmeca ProX technical manual . information see 5. Press the SELECT key briefly. The mA display will start to flash.
The imaging mode can be changed by pressing the MODE key briefly. 3. Press the SELECT key briefly three times. The density value can now be changed with the parameter adjustment keys. User's manual Planmeca ProX...
Steps 1. Select the large or small animal mode of which the preprogrammed settings is to be programmed. The indicator light of the selected mode will come on. Planmeca ProX User's manual...
Page 31
The time display and the ready light will start to flash. The imaging mode can be changed by pressing the MODE key briefly. 4. Modify the exposure time. The exposure time value is changed with the parameter adjustment keys. User's manual Planmeca ProX...
Page 32
You will hear a signal tone. NOTE If you interrupt programming for over 45 seconds, the unit automatically exits the programming mode, and the current values will be stored in the memory. Results The new preprogrammed settings are set. Planmeca ProX User's manual...
± 15VDC voltage is out of limits. Er.61 Communication error between control panel and tube head CPU. Er.71 FLASH memory check-sum error (tube head CPU). Er.81 EEPROM memory defective (tube head CPU). Er.83 Config register error (tube head CPU). User's manual Planmeca ProX...
The unit surfaces can be cleaned with a soft cloth damped in a mild cleaning solution. Stronger agents can be used for disinfecting the surfaces. We recommend Dürr System-hygiene FD 322 or respective disinfecting solution. Planmeca approved cleaning agents Manufacturer Brand name Alpro Medical...
To guarantee user and patient safety and to ensure image quality the unit must be checked and recalibrated by a qualified PLANMECA service technician once a year or after every 10 000 exposures if this is sooner. Please refer to the Planmeca ProX technical manual for complete servicing information. User's manual...
1. Secure the support arm to the transport position with strap. CAUTION Never move the Planmeca ProX mobile x-ray unit without first securing the support arm to the transport position. 2. Disconnect the Planmeca ProX from the power supply and loop the cable around the handle. User's manual Planmeca ProX...
Page 38
Not moving from the handle might cause the mobile unit to overbalance. NOTE If you have to move the ProX Mobile X-ray unit over a low obstacle, tilt the base slightly by pressing the back of the wheel assembly with your foot.
(6.3x32mm) (special fuse, manufacturer Bussmann, type MDA) Units with 220-240V~ voltage setting: 8AT, 250V, slow blow (6.3x32mm) (special fuse, manufacturer Bussmann, type MDA) Duty cycle 1:13,5 automatic control, at least 6 s automatic control User's manual Planmeca ProX...
Page 40
No other equipment should be connected to the same fused mains line as the X-ray unit. In some countries an additional external fault current guard is also required. Original manufacturer PLANMECA OY Asentajankatu 6 FIN-00880 Helsinki FINLAND Phone: +358-20-7795 500 Planmeca ProX User's manual...
18 Technical specifications 18.4 Installation options Standard wall mount or remote control panel 1 Standard wall mount 2 Remote control panel Fixed control panel with double exposure button Planmeca ProX User's manual...
18 Technical specifications 18.5 User's statement for Planmeca ProX Radiation leakage technique factors The maximum rated peak tube potential is 70 kV and the maximum rated continuous tube current is 0.14 mA for the maximum rated peak tube potential. Minimum filtration The radiation port contains an added 1.5 mm aluminium filtration.
Page 48
The nominal electric power for a load time of 0.1 sec and at the nominal x- ray tube voltage 1.4 kW at 70 kV, 8 mA Anode heating/cooling curve of the X-ray tube X-ray tube assembly heating/cooling curve Planmeca ProX User's manual...
Page 49
3.1 kg Values of loading factors concerning leakage radiation 70 kV, 8 mA Tolerances of the focal spot on the reference axis X= ±0.5 mm (sideways) Y= ±0.5 mm (in depth) Z= ±0.5 mm (in height) User's manual Planmeca ProX...