Fusing
Duty cycle
Electrical classification
Mechanical data
Weight
Color
Environmental requirements
Temperature
Humidity
Atmospheric pressure range
External mains fuse recommendation
The recommendation for the external mains fuses are:
•
units with 100V~ or 115V~ voltage setting:
•
units with 220-240V~ voltage setting:
No other equipment should be connected to the same fused mains line as the x-ray unit. In
some countries an additional external fault current guard is also required.
Original manufacturer
PLANMECA OY
Asentajankatu 6
FIN-00880
Helsinki
FINLAND
Phone: +358-20-7795 500
User's manual
22 TECHNICAL SPECIFICATIONS
units with 100V~ or 110-115V~ voltage
setting: 15AT, 250V, slow blow (6.3x32mm)
(special fuse, manufacturer Bussmann,
type MDA)
units with 220-240V~ voltage setting:
8AT, 250V, slow blow (6.3x32mm)
(special fuse, manufacturer Bussmann, type MDA)
1:13,5 automatic control, at least 6 s automatic
control
Class I Type B
total 33 kg (73 lbs)
tube head
4.2 kg (9.3 lbs) with standard cone
4.5 kg (10 lbs) with long cone
RAL 9016
operating +5°C - +40°C
storage -10°C - +50°C
transport -10°C - +50°C
25% - 75%
700 hPa - 1060 hPa
16A, time lag
10A, time lag
Planmeca ProX
43