TRIGger:A:PULse:RUNT:WIDth
Description
This command sets or queries the minimum width for a valid A pulse runt trigger. This command is
equivalent to selecting Runt Setup from the Trig menu and then setting the desired Width.
Group
Trigger
Related Commands
TRIGger:A:PULse:RUNT:WHEn (see page 392)
Syntax 1
TRIGger:A:PULse:RUNT:WIDth <NR3>
Syntax 2
TRIGger:A:PULse:RUNT:WIDth?
Argument
• <NR3>
This specifies the minimum width, in seconds.
Example 1
TRIGger:A:PULse:RUNT:WIDth?
This query might return
, indicating that the minimum width
:TRIGGER:A:PULSE:RUNT:WIDTH 2.0000E-09
of a pulse runt trigger is 2 ns.
Example 2
TRIGger:A:PULse:RUNT:WIDth 15E-6
This command sets the minimum width of the pulse runt trigger to 15 µs.