*ESE
Description
This command sets and queries the bits in the Event Status Enable Register (ESER). The ESER
prevents events from being reported to the Status Byte Register (STB). For a more detailed discussion of
the use of these registers, see Registers on page 522.
Group
Status and Error
Related Commands
*CLS (see page 318), DESE (see page 331), *ESR? (see page 320), EVENT? (see page 332), EVMsg?
(see page 333), *SRE (see page 326), *STB? (see page 327)
Syntax 1
*ESE <NR1>
Syntax 2
*ESE?
Argument
• <NR1>
This specifies the binary bits of the ESER according to this value, which ranges from 0 through 255.
The power-on default for ESER is 0 if *PSC is 1. If *PSC is 0, the ESER maintains the previous power
cycle value through the current power cycle.
Note: Setting the DESER and the ESER to the same values allows only those codes to be entered into
the Event Queue and summarized on the ESB bit (bit 5) of the Status Byte Register. Use the DESE
command to set the DESER.
Example 1
*ESE 209
This command sets the ESER to binary 11010001, which enables the PON, URQ, EXE and OPC bits.
Example 2
*ESE?
This query form of this command returns
10111010.
, showing that the ESER contains the binary value
186