TRIGger:B:BY
Description
This command selects or returns whether the B trigger occurs after a specified number of events or a
specified period of time after the A trigger. This is equivalent to selecting B Event (Delayed) Trigger Setup
from the Trig menu, selecting the A→B Seq tab, and then choosing Trig After Time or Trig on nth event.
The traditional Runs After functionality is now served by the Horizontal Delay function. For
Note:
details, see the HORIZontal[:MAIN]:DELay:MODe and HORIZontal[:MAIN]:DELay:TIMe commands.
Group
Trigger
Related Commands
TRIGger:B:EVENTS:COUNt (see page 420), TRIGger:B:TIMe (see page 423),
HORIZontal[:MAIN]:DELay:MODe (see page 184), HORIZontal[:MAIN]:DELay:TIMe (see page 185)
Syntax 1
TRIGger:B:BY {EVENTS|TIMe}
Syntax 2
TRIGger:B:BY?
Arguments
• EVENTS
This sets the B trigger to take place following a set number of trigger events after the A trigger occurs.
The number of events is specified by TRIGger:B:EVENTS:COUNt.
• TIMe
This sets the B trigger to be ready to occur a set time after the A trigger event. The time period is
specified by TRIGger:B:TIMe.
Example 1
TRIGger:B:BY?
This query might return
:TRIGGER:B:BY EVENTS
number of trigger events after the A trigger occurs.
Example 2
TRIGger:B:BY TIMe
This command sets the B trigger to be ready to occur a set time after the A trigger event.
414
, indicating that the B trigger takes place following a set