MEASUrement:IMMed:VALue?
Description
This query-only command returns the value of the measurement specified by the
MEASUrement:IMMed:TYPe command. The measurement is immediately taken on the source(s)
specified by a MEASUrement:IMMed:SOURCE command.
A change to HORizontal:MAIn:SCALe or CH<x>:SCALe will not necessarily have taken affect if
Note:
followed by this command.
Group
Measurement
Related Commands
MEASUrement:IMMed:TYPe (see page 236), MEASUrement:IMMed:SOURCE[1] (see page 234),
MEASUrement:IMMed:SOURCE2 (see page 235), *ESR? (see page 320), ALLEV? (see page 329)
Syntax
MEASUrement:IMMed:VALue?
Example
MEASUrement:IMMed:VALue?
This query might return
:MEASUREMENT:IMMED:VALUE 9.9000E+37
warning associated with it, then an item is added to the error queue. The error can be checked for with
the *ESR? and ALLEV? commands.
. If the measurement has an error or