Cleaning And Maintenance; Cleaning; Servicing - Sartorius PMA.Vision Operating Instruction

Paint-mixing scale for use in potentially explosive atmospheres of class i, zone 1 and class i, division 1
Hide thumbs Also See for PMA.Vision:
Table of Contents

Advertisement

Cleaning and Maintenance

VIS1X
48
Operating Instructions
10 Cleaning and Maintenance

10.1 Cleaning

Before cleaning the power supply, ex-link converter, or the scale: Disconnect all
devices from the power supply.
WARNING Electrical Hazard from Voltage or Current
Disconnect the power supply (if connected) from the mains. Unplug any connected
data cables from the ex-link converter. Never open the scale or the power supply.
The parts contained in these cannot be cleaned, repaired or replaced by the operator.
NOTICE
Do not clean the following parts with acetone or aggressive cleaning agents:
− Mains socket
− Data interface
− Labels, and all other plastic parts
Procedure
Disconnect the device from the power supply.
t
NOTICE Make sure that no liquid or dust gets into the scale or the power supply.
t
NOTICE Corrosion or damage to the device due to unsuitable cleaning agents!
t
Do not use corrosive, chloride-containing and aggressive cleaning agents.
t
Do not use cleaning agents that contain abrasive ingredients, e.g. scouring
t
agents, steel wool.
Only use soft brushes and cloths for cleaning.
t
Do not use solvent-based cleaning agents.
t
Cleaning the Control Panel
Before cleaning the control panel: Turn off the device as touching the screen
t
could trigger unwanted inputs.
Cleaning the Device Housing
Wipe off the housing with a slightly damp cloth. In the event of more severe
t
contamination, use a mild soap solution.
Wipe the device with a soft cloth.
t

10.2 Servicing

To ensure the continued accuracy of your scale, we recommend scheduling regular
servicing at least once a year. Sartorius Service offers different service contracts with
maintenance intervals that are tailored to your needs.
A calibration certificate should always be issued as part of every maintenance
session. Safety inspections of the power supply and its connections must be
performed at appropriate intervals by a qualified electrician (e.g. every two years).