Electronic paint-mixing scales for use in zone 1 hazardous areas

Advertisement

Quick Links

Installation Instructions
Sartorius PMA.World | PMA.Net
Model PMA 7501-X00G | PMA 7501-X00GL
Electronic Paint-mixing Scales for Use in Zone 1 Hazardous Areas
98648-015-22

Advertisement

Table of Contents
loading

Summary of Contents for Sartorius PMA.World

  • Page 1 Installation Instructions Sartorius PMA.World | PMA.Net Model PMA 7501-X00G | PMA 7501-X00GL Electronic Paint-mixing Scales for Use in Zone 1 Hazardous Areas 98648-015-22...
  • Page 2: General View Of The Equipment

    General View of the Equipment PMA7501-X00G | PMA7501-X00GL (PMA.World | PMA.Net) Electronic Paint-mixing Scales for Use in Zone 1 Hazardous Areas 1 Display and control unit 8 c key (Clear) and [REC] key for 2 Q key (On/Standby) paint-mixing applications...
  • Page 3: Table Of Contents

    Documents ..... 27 ufacturers’ paint-mixing applications) installed on your PC. Communication is carried out via the data interface. If required, Sartorius can supply you with DOS- or WINDOWS ® -based device drivers to generate application programs.
  • Page 4: Pma7501-X00G | -X00Gl

    Make sure the equip- the installation. If you need assistance, ment is accessible at all times. contact your Sartorius dealer or the Sar- torius Service Center. General Provisions for Installing – Avoid static electricity. Connect an equi- the PMA7501-X…...
  • Page 5 Sartorius equipment, and disconnect any prevent build-up of excessive heat. wires that are assigned differently. The – Use original Sartorius spare parts, only! operator shall be solely responsible for any damage or injuries that occur when – Never use a hammer to close the lid of a using cables not supplied by Sartorius.
  • Page 6: Getting Started

    Getting Started § Remove the scale from its packaging. § After unpacking the scale, check it immediately for any visible damage as a result of rough handling during shipment. Equipment Supplied for Model PMA7501-X00G – Scale – Weighing pan – Power supply –...
  • Page 7 If you should have any questions, please contact your supplier or Sartorius Customer Service for information on the legal regulations applicable in your country. Use only original Sartorius power supplies. The use of power supplies from other manufacturers, even if these units have a registered approval rating from a national testing laboratory, requires the approval of a certified technician.
  • Page 8 > If the YC011-Z EX-link converter is connected, LED on the converter lights up to indicate that it is powered. Connecting Data Cables Important Note: Connect/disconnect data cables to or from equip- ment used in the potentially explosive area Zone 1 only after the power supply has been disconnected.
  • Page 9 Data cable connection only for PMA7501-X00GL (Ethernet interface RJ45) § Plug the data cable of the EX link converter YCO11-Z into the Ethernet interface port RJ45. § Check the switch position (16). The switch must be set to the left position (Ethernet interface).
  • Page 10: Operation

    Operation Turn on the scale using the After the scale has been If a different readout is Q key (2). turned on, it will automati- displayed, zero or tare the cally run a self-test. scale using the U key (5). At the end of this test, 0.0 g is displayed.
  • Page 11 Weighing with Two Decimal Places Note: To weigh using two decimal places, you must first adapt the settings (refer to the chapter entitled “Menu Settings”) Press the I key (6). The display shows “0.00 g.” Place an empty paint can on the weighing pan (11). Press the U key (5).
  • Page 12: Applications

    Applications Formulation Mode (Calculation by a Factor) This mode enables you to weigh in amounts that are smaller or larger than that of your basic formula for a specific paint color (e.g., 250 ml of a 1-l formula). You can select various factors (amounts) by pressing the K formulation key (7): 0.25 0.5 0.75 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0.
  • Page 13 3. “.25” is displayed next to the weight 4. Slowly pour in the first component, “250 g”, until the display shows “250 g.” 5. Pour in the second component, “250 g”, until the display shows “500 g.” 6. Pour in the last component, “500 g”, until “1000 g”...
  • Page 14 Example (cumulative): Place an empty paint can on Press the U key (5) Pour in the first component. the weighing pan (11). 0.0 g + 50.0 g + 118.0 g Press the L key [MEM] (9). Pour in the 2nd component. Press the L key [MEM] (9).
  • Page 15: Calibration/Adjustment

    Calibration/Adjustment You can calibrate/adjust the scale by pressing the U key (5). Calibration weight: 5,000 g; accuracy: + 0.075 g. After connection to AC power and before each calibration/adjustment, allow the scale to warm up for approx. 30 min. Hold down the U key (5) for 2 sec. When 5000 is displayed, release the key.
  • Page 16: Menu Settings

    § Press the c key (Clear) several times to exit the menu. Note: To obtain a detailed list of the menu codes, please ask your nearest Sartorius office! Please note that English screens are available for the scale even if they are shown in German.
  • Page 17 Important Menu Settings § Hold down the L key [ENTER] for approx. 2 sec. “SETUP” will appear on the display (Level 1). Level 1 SETUP Language Settings Level 1 Level 2 Level 3 Level 4 § Press the R key to select “LANGUAGE” LANGUAGE §...
  • Page 18 Assigning the I Toggle Key (6) By pressing the toggle key I (6), the scale toggles between the basic setting (see page 14) and the settings defined under “SETUP- APPLICATION – UNIT” and “SETUP- APPLICATION – DECIMALS” Level 1 Level 2 Level 3 Level 4 §...
  • Page 19 Note: Keep the numeric code in a safe place! The scale can only be accessed by entering the correct code! Level 1 Level 2 Level 3 Level 4 § Press the RS keys to select “INPUT” INPUT § Press the L key [ENTER] PASSWORD §...
  • Page 20 RS232 EXT. § Press the L key [ENTER], “o” will appear, indicating that the desired interface is activated RS232 inT. § Press the c key (Clear) several times to exit the menu Note: To obtain a detailed list of the menu codes, please ask your nearest Sartorius office.
  • Page 21: Troubleshooting

    Troubleshooting Guide Problem Cause Solution o segments appear – No AC power available – Check the AC power supply on the weight display Weight display shows – The weighing pan is – Position the weighing pan “Low” is not in place Weight display shows –...
  • Page 22: Care And Maintenance

    Information shipment. $ Storage temperature: –20°C to +75°C In this case, notify your nearest Sartorius $ Permissible moisture level for storage of Service Center or the International Tech- nical Support Unit based in Goettingen, the packaged scale: 90% max.
  • Page 23: Recycling

    EU: 1) Go to http://www.sartorius.com. 2) Select the summary under “Service.” 3) Then select “Information on Disposal.” 4) Addresses for local Sartorius disposal contacts can be found in the PDF files given on this webpage.
  • Page 24: Interface Description

    Interface Description Internal data interface RS-232 (PMA.7501-X00G | PMA.7501-X00GL) Pin assignment Data output 9-contact RS-232 data cable Pin 2: (RXD) Receive Data Pin 3: (TXD) Transmit Data Pin 4: (DTR) Data Terminal Ready Pin 5: (GND) Ground Pin 6: BPI bridge Pin 8: (CTS) Clear to Send Format:...
  • Page 25: Specifications

    Specifications Model PMA7501-X00G | -X00GL eighing range 999.95/7500 Readability 0.05/0.1 Tare range (subtractive) –999.95/–7500 Max. linearity <±0.2 Stability range digit 0.25 to 4 Moisture-proof rating Non-condensing Allowable ambient operating °C 0...+40 temperature range Weighing pan d mm Scale housing (W Net weight, approx.
  • Page 26: Accessories

    Accessories In-use dust cover YDC01PMA EX power supplies (PMA7501-X00G) EU/UK YPS05-ZDP USA/CDN YPS05-ZKP + country-specific power cord EX power supplies (PMA7501-X00GL) EX-link converter YC011-Z +country-specific power cord Cables for connection to YCO11-Z EX-link converter May be used only outside the potentially explosive atmosphere. RS-232 data cable (SBI) (2 m) YCC01-0027M2...
  • Page 27: Documents

    Documents...
  • Page 28 28 28...
  • Page 38 1.61 W 575 nF 10 µF (for IIB) 420 µH 5 µH (for IIB) Non-hazardous area Sartorius cable; permanently Sartorius cable; permanently Hazardous area mounted on the power supply/ mounted on the power supply Zone 1 Ex-link box (max. length: 100m) (max.
  • Page 39 in YDI02-Z.. YDI05-Z.. Z966 YCO01-Y 12.4 V 12.6 V 12 V 11.8 V 12.6 V 24.8 V 25.2 V 23.6V 25.2 V 24 V 260 mA * 123 mA * 131mA 328 mA * 800 mW * 361 mW * 0.96 W * 1.24 µF 1.5 µF...
  • Page 40 Ethernet or RS-232 data YCO11-Z interface EX-link box for connecting to equipment using 250V II (2)G [EEx ib] IIB power supply, maximum Sartorius cable; permanently mounted on the Ex-link box (max. length: 100m) Non-hazardous area Hazardous area Zone 1 Gas: IIB T4 TM01-X...
  • Page 41 Sartorius Lab Instruments GmbH & Co. KG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone: +49.551.308.0 Fax: +49.551.308.3289 www.sartorius.com The information and figures contained in these instructions correspond to the version date specified below. Sartorius reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.