Establishing The Power Supply - Sartorius PMA.Vision Operating Instruction

Paint-mixing scale for use in potentially explosive atmospheres of class i, zone 1 and class i, division 1
Hide thumbs Also See for PMA.Vision:
Table of Contents

Advertisement

Connect the cable lug of the equipotential bonding cable to the grounding
t
terminal of the scale.
Connect the equipotential bonding cable to the customer-supplied equipotential
t
bonding conductor.
Connect the ex-link converter to the equipotential bonding conductor using another
equipotential bonding cable with a gage of at least 4 mm
Connect the cable lug of the equipotential bonding cable to the grounding
t
terminal of the ex-link converter.
Connect the equipotential bonding cable to the customer-supplied equipotential
t
bonding conductor.
Connecting a PC/Notebook
Plug an Ethernet cable (1) into the Ex-Link converter and connect the cable to a
t
Windows PC (direct connection) or to a network.
1
5.6

Establishing the Power Supply

NOTICE Make sure that the voltage rating printed on the power supply is identical
to your local supply voltage. The scale is connected to the power supply using the
power supply YPS07-USB (see Chapter "16 Accessories", page 56), which is
supplied with mains adapters for use in various countries. If the stated supply
voltage or the plug design of the power supply does not comply with your country's
standard then please inform your nearest Sartorius representative.
Power supply assembly is described in the following.
Installation
.