Sartorius PMA.Quality PMA7501H Installation Instructions Manual
Sartorius PMA.Quality PMA7501H Installation Instructions Manual

Sartorius PMA.Quality PMA7501H Installation Instructions Manual

Electronic paint-mixing scales

Advertisement

Installation Instructions
Sartorius PMA.Quality
PMA7501H
Electronic Paint-mixing Scales
98648-014-35

Advertisement

Table of Contents
loading

Summary of Contents for Sartorius PMA.Quality PMA7501H

  • Page 1 Installation Instructions Sartorius PMA.Quality PMA7501H Electronic Paint-mixing Scales 98648-014-35...
  • Page 2: General View Of The Equipment Pma7501H

    General View of the Equipment PMA7501H 1 Display and control unit 10 Display 2 Q key (On/Standby) 11 Weighing pan 3 R key: Upwards 12 Interfaces (D-Sub plug, 9-contact) 4 S key: Downwards 13 Connection to AC power 5 U key: Zero/Tare 14 Column 6 I key (TOGGLE) 15 Joint...
  • Page 3: Table Of Contents

    Getting Started ....5 tions) installed on your PC. Operating the PMA7501H ... . 7 If required, Sartorius can supply you with ® DOS or Windows -based device drivers to Applications .
  • Page 4: Warnings And Safety Precautions

    Improper use questions, please contact your supplier or or handling can result in property damage Sartorius Customer Service for informa- and/or personal injury. To prevent damage tion on the legal regulations applicable in to the equipment, read these operating your country.
  • Page 5: Getting Started

    Getting Started § Remove the scale from its packaging. § After unpacking the scale, check it immediately for any visible damage as a result of rough handling during shipment. Equipment Supplied – Scale – Weighing pan – AC adapter or power supply –...
  • Page 6 If it does not match your local voltage rating or if the plug design differs from your local standard, contact your Sartorius office or dealer. Use only genuine Sartorius AC adapters. The use of AC adapters from other manufacturers,...
  • Page 7: Operating The Pma7501H

    Operating the PMA7501H Turn on the scale using the After the scale has been If a different readout is Q key (2). turned on, it will automati- displayed, zero or tare the cally run a self-test. scale using the U key (5). At the end of this test, 0.0 g is displayed.
  • Page 8 Weighing with Two Decimal Places Note: To weigh using two decimal places, you must first adapt the settings (refer to the chapter entitled “Menu Settings”) Press the I key (6). The display shows “0.00 g.” Place an empty paint can on the weighing pan (11). Press the U key (5).
  • Page 9: Applications

    Applications Formulation Mode (Calculation by a Factor) This mode enables you to weigh in amounts that are smaller or larger than that of your basic formula for a specific paint color (e.g., 250 ml of a 1-l formula). You can select various factors (amounts) by pressing the K formulation key (7): 0.25 0.5 0.75 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0.
  • Page 10 3. “.25” is displayed next to the weight 4. Slowly pour in the first component, “250 g” of green paint, until the display shows “250 g.” 5. Pour in the second component, “250 g” of red paint, until the display shows “500 g.” 6.
  • Page 11 Example: Place an empty paint can on Press the U key (5) Pour in the first component. the weighing pan (11). 0.0 g + 50.0 g + 118.0 g Press the L key [MEM] (9). Pour in the 2nd component. Press the L key [MEM] (9).
  • Page 12: Calibration/Adjustment

    Calibration/Adjustment You can calibrate/adjust the scale by pressing the U key (5). To do so, menu code 1 5 1 must be selected. Calibration weight: 5,000 g; accuracy: + 0.075 g. After connection to AC power and before each calibration/adjustment, allow the scale to warm up for approx.
  • Page 13: Menu Settings

    § Press the L key [ENTER]. “o” will appear. The new code is stored. § Press the c key (Clear) several times to exit the menu. Note: To obtain a detailed list of the menu codes, please ask your nearest Sartorius office.
  • Page 14 Important Menu Settings § Hold down the L key [ENTER] for approx. 2 sec. “SETUP” will appear on the display (Level 1). Level 1 SETUP Language Settings Level 1 Level 2 Level 3 Level 4 § R key: select “LANGUAGE” LANGUAGE §...
  • Page 15 Toggling Decimal Places (Standard = 1 decimal place PolyRange = 2 decimal places) Toggling units (Grams or PT./PD.) These settings are active when the scale is switched on. Level 1 Level 2 Level 3 Level 4 § Press the L key [ENTER] SETUP §...
  • Page 16 Note: Keep the numeric code in a safe place! The scale can only be accessed by entering the correct code! Level 1 Level 2 Level 3 Level 4 § RS keys: select “INPUT” INPUT § Press the L key [ENTER] PASSWORD §...
  • Page 17 § Press the L key [ENTER]; “o” will appear: etc. the new code is set § Press the c key (Clear) several times o CODES to exit the menu. Note: To obtain a detailed list of the menu codes, please ask your nearest Sartorius office.
  • Page 18: Troubleshooting Guide

    Troubleshooting Guide Problem Cause Solution No segments appear – No AC power available – Check the AC power supply on the weight display Weight display shows – The weighing pan is – Position the weighing pan “Low” not in place Weight display shows –...
  • Page 19: Care And Maintenance

    $ Read and follow the instructions given in § Observe the Warning and Safety the section entitled “Safety Inspection.” Information In this case, notify your nearest Sartorius Service Center or the International Tech- nical Support Unit based in Goettingen, Germany. Maintenance and repair work...
  • Page 20: Recycling

    EU: 1) Go to http://www.sartorius.com. 2) Select the summary under “Service.” 3) Then select “Information on Disposal.” 4) Addresses for local Sartorius disposal contacts can be found in the PDF files given on this webpage.
  • Page 21: Interface Port

    Interface Port Data Interface Pin Assignment 9-contact interface port Pin 2: (RXD) Receive Data Pin 3: (TXD) Transmit Data Pin 4: (DTR) Data Terminal Ready Pin 5: (GND) Ground Pin 6: BPI bridge Pin 8: (CTS) Clear to Send Specifications Model PMA7501H Weighing range...
  • Page 22: Accessories

    Accessories Order no.: YDC01PMA In-use dust cover YCC01-0027M2 RS-232 Data cable (SBI) (2 m) YCC01-0028M2 RS-232 Data cable (BPI) (2 m) YCC01-0028M20 RS-232 Data cable (BPI) (20 m) YCO12 USB/RS-232 Data cable (SBI) (1.80 m) YCO13 USB/RS-232 Data cable (BPI) (1.80 m) AC adapter 6971991...
  • Page 24 Sartorius Lab Instruments GmbH & Co. KG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 +49.551.308.32 89 www.sartorius.com Copyright by Sartorius Lab Instruments GmbH & Co. KG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius.