Cleaning and Maintenance

1
76
Cubis
®
MCA Operating Instructions
9
Cleaning and Maintenance
9.1
Preparing a Device with an Analytical Draft Shield or
Flat Glass Draft Shield
Procedure
Turn the device off .
t
Disconnect the device from the power supply . To do so, disconnect the power
t
supply cable from the wall outlet .
Fully open the draft shield side panels and upper panel .
t
Press and hold the locking button (1) on the door handle and pull the upper
t
panel completely out of the guide rails .
If this relates to a device with an analytical draft shield: Press and hold the
t
1
locking button (1) on the guide rails and pull the side panels completely out of
the guide rails .
If this relates to a device with a flat glass draft shield: Press and hold the locking
t
button (1) on the door handle and pull the side panels completely out of the
guide rails .
1