MODE
2) Press key
3) Wait till
appears, put weights of 1/3 of full load on the weighing pan and press
TARE
down key
4) Wait till
appears, and then put weights of 2/3 of full load on pan and press down
TARE
key
E.g., for JWE-3K, 2/3 of full load is 2kg.
5) Wait till
appears, then put weight of full load on and press
JWE-3K, full load is 3kg.
6) The calibration procedure ends up with a symbol of "PASS".
7) Remove all the weights, and then press key
7. Setting Mode& Description of Parameter Values
Setting Mode
MODE
1) Long-press key
displays "CAL", then release the keys.
MODE
2) Press key
M+MR
3) Press key
MC/SET
4) Press key
setting and back to the weighing mode. If more settings are needed, press key
Description of Parameter Values
1)
Calibration See "7. Calibration"
2)
Set the active weight unit. The following setting are available:
kg, g, lb, lb-oz,斤, 台斤, 港斤,pcs and % .
3)
Set the init weight unit. The following setting are available:
kg, g, lb, lb-oz,斤, 台斤, 港斤,pcs and %.
to enter the zero point calibration mode.
E.g., for JWE-3K ,1/3 of full load is 1kg.
while powering on the scale to enter setting mode. The screen
to circle through setting modes.
to choose function.
switch on and off the chosen function. Press key
ZERO+
to return to weighing mode.
(Enable);
(Enable);
JWE
TARE
E.g., for
ZERO+
to save the
MODE
(disable)
(Disable)
again.