Device Locking; Atmospheric Pressure Compensation - Ahlborn MA2790BTFV Operating Instructions Manual

Wireless sensor link ma2790btfv with almemo 2790 and sensor module
Table of Contents

Advertisement

13. Device configuration
In sleep mode the function field displays ab-
breviation ´SLP´; in this status the device can
no longer be operated.
To quit sleep mode and terminate a measu-
ring operation the device must be switched
ON again by pressing key

13.2 Device locking

Measured data acquisition can be locked by
means of a password and thus protected
against unauthorized manipulation. To do so
- before starting a measuring operation (see
13.1) - select the locking code function ´VC´
and enter a 4-digit number as password (dis-
play ´Loc´).
When the measuring operation is started all
keys are locked - except 'ON'. If key
pressed the ´VC´ function will reappear and, as soon as the appropriate num-
ber is entered, all keys will be unlocked again. Otherwise the measuring ope-
ration is simply continued.

13.3 Atmospheric pressure compensation

Some measured variables depend on the ambient atmospheric pressure (see
11.1.2, measuring range list 'with PC') - with the effect that large deviations
from normal pressure (1013 mbar) may lead to measuring errors.
e.g. Error per 100 mbar
RH, psychrometric
Mixture ratio
Dynamic pressure
O
saturation
2
It is important therefore, especially when working at significant altitudes above
sea level, to take due account of atmospheric pressure (approx. -11 mbar / 100
meters above mean sea level, MSL).
The atmospheric pressure can be entered as
a parameter in function 'mb'. (see 13) With op-
tion RHA (integrated atmospheric pressure
sensor) the current atmospheric pressure is
displayed in this function and used automati-
cally for all compensation procedures. There
is also already a function channel 'AP' with the
appropriate measured value. (see 9.2)
24
ALMEMO
.
ON
approx. 2%
approx. 10 %
approx. 5%
approx. 10%
2790
®
is
ON
Compensation range
500 to 1500 mbar
Vapor pressure VP up to 8 bar
800 to 1250 mbar (error <2%)
500 to 1500 mbar

Advertisement

Table of Contents
loading

This manual is also suitable for:

Almemo 2790

Table of Contents