Atmospheric Pressure Compensation - Ahlborn ALMEMO 2490-1A Operating Instructions Manual

Universal measuring instruments
Table of Contents

Advertisement

different ways or even switched off in cases where there is a risk of accidentally
activating relative measuring by zero-setting the measured value.
Loc parameter :
0 The offset is saved in RAM, base or zero-point - depending on locking
1 The offset is saved in RAM only.
2 Relative measuring is locked

Atmospheric pressure compensation

12.5
Measured variables dependent on the ambient atmospheric pressure (see Man-
ual 6.3.3 Measuring range list ´with PC´) may, in the event of large deviations
from normal pressure (1013 mbar), involve certain measuring errors.
e.g. Error per 100 mbar
Rel.humidity psychrometer
Mixture ratio, capacitive
Dynamic pressure
O
saturation
2
It is advisable therefore, especially when taking measurements at appreciable
heights above sea level to take due account of the atmospheric pressure (ap-
prox. -11 mbar / 100 meters above mean sea level, MSL).
The appropriate atmospheric pressure can be entered in parameter 'mb'either
in device programming or it can be measured using an atmospheric pressure
sensor (reference sensor with designation ´*P´, see Manual 6.7.2).
Compensation range
approx. 2% 500 to 1500 mbar
approx. 10 % Vapor pressure VP up to 8 bar
approx. 5% 800 to 1250 mbar (error < 2%)
approx. 10% 500 to 1500 mbar
12. Device configuration
29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Almemo 2490-2a

Table of Contents